Class IIReported 2026-01-07
BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
Recalled by Beaver-Visitec International, Inc.
Reason for Recall
Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended
Advertisement
The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).