MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Recalled by REPLIGEN CORPORATION

Reason for Recall

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact

Advertisement

At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

Recall Number
Z-1024-2026
Classification
Class II
Status
Ongoing
Category
device
Initiated
2025-12-10
Reported
2026-01-21
Distribution
U.S.
Code Info
PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496