MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Recalled by Elekta, Inc.

Reason for Recall

Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

Advertisement
Recall Number
Z-1078-2026
Classification
Class II
Status
Ongoing
Category
device
Initiated
2025-12-10
Reported
2026-01-21
Distribution
US Nationwide distribution in the state of TX.
Code Info
MOSAIQ v 3.10.200, and 3.1.1.0 up to and including v 3.2.3.1 with Particle Base. Software Versions: 3.10.200, 3.1.1.0, 3.1.2.0, 3.1.3.0, 3.1.3.1, 3.1.3.2, 3.1.3.3, 3.1.3.4, 3.2.0.0, 3.2.1.0, 3.2.1.1, 3.2.1.2, 3.2.1.3, 3.2.1.4, 3.2.2.0, 3.2.2.1, 3.2.3.0, 3.2.3.1 UDI-DI: 07340201500026, 07340201500071