WhichRecall
Class IIReported 2026-01-21

Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable

Recalled by C-RAD POSITIONING AB

Reason for Recall

Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups

Advertisement

n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.

Recall Number
Z-1092-2026
Classification
Class II
Status
Ongoing
Category
device
Initiated
2025-10-09
Reported
2026-01-21
Distribution
The software module cSRS has been installed in 37 countries Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, CA, CT, FL, MO, NC, NM, NY, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Australia, Austria, Belarus, Belgium, Benin, Canada, China, Colombia, Egypt, El Salvador, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Mexico, Morocco, Netherlands, New Zealand, Norway, Portugal, Romania, Russia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Ukraine, and Vietnam.
Code Info
Lot Code: Note: The error is not related to any serial numbers or LOT numbers. The error is limited to the software module cSRS in combination with cMotion module
Source: openFDA enforcement report. WhichRecall is not affiliated with the FDA. See our disclaimer.
Advertisement