ICU Medical Plum Duo Precision IV Pump, List Numbers: 400020403 (US), 400021003 (Philippines)

Recalled by ICU Medical, Inc.

Reason for Recall

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy

Advertisement

The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

Recall Number
Z-1119-2026
Classification
Class II
Status
Ongoing
Category
device
Initiated
2025-12-15
Reported
2026-01-28
Distribution
Worldwide distribution - US Nationwide and the country of Philippines.
Code Info
400020403 (US), Software Versions: 1.2.2; 400021003 (Philippines), Software Versions: 1.2.0