WhichRecall
Class IIReported 2026-02-04

Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 10041 Conductivity Standard Solutions are a secondary standard solution used for the calibration of conductivity cells together with conductivity meters

Recalled by NIPRO Technical Services, Inc.

Reason for Recall

due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.

Advertisement
Recall Number
Z-1185-2026
Classification
Class II
Status
Ongoing
Category
device
Initiated
2025-12-12
Reported
2026-02-04
Distribution
U.S. Nationwide distribution in the states of AL, AR, AZ, CA, DE, FL, GA. IL, IN, KS, MI, MO, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN TX, UT, VA, and WA.
Code Info
Catalog/Model Numbers/ Lot Numbers / UDI codes: 10001 14250616-1019106 (01) 0 0850016 97600 0 (17) 251030 (10) SHOWCASE 10011 14250616-1019107 (01) 0 0850016 97609 3 (17) 251030 (10) SHOWCASE 10040c 50250624-1019120 (01) 0 0850016 97636 9 (17) 251030 (10) SHOWCASE 10040 50250624-1019119 (01) 0 0850016 97635 2 (17) 251030 (10) SHOWCASE 10032 14250616-1019108 (01) 0 0850016 97631 1 (17) 251030 (10) SHOWCASE 10041 50250522-1018643 (01) 0 0850016 97637 6 (17) 251030 (10) SHOWCASE 10041 50250624-1019121 (01) 0 0850016 97637 6 (17) 251030 (10) SHOWCASE 10021 15F250603-1018834 (01) 0 0850016 97619 2 (17) 251030 (10) SHOWCASE
Source: openFDA enforcement report. WhichRecall is not affiliated with the FDA. See our disclaimer.
Advertisement