IQon Spectral CT; Product Code (REF): 728332;

Recalled by Philips North America Llc

Reason for Recall

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry

Advertisement

One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

Recall Number
Z-1316-2026
Classification
Class II
Status
Ongoing
Category
device
Initiated
2026-01-07
Reported
2026-02-18
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Macao, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, R¿union, Russian Fed., Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkmenistan, United Kingdom, Uzbekistan, Vietnam.
Code Info
Product Code (REF): 728332; UDI-DI: 00884838059542; Serial Numbers: 60128, 60141, 60151, 60161, 60171, 60181, 60129, 60142, 60152, 60162, 60172, 60182, 60130, 60143, 60153, 60163, 60173, 60132, 60144, 60154, 60164, 60174, 60133, 60145, 60155, 60165, 60175, 60134, 60146, 60156, 60166, 60176, 60136, 60147, 60157, 60167, 60177, 60137, 60148, 60158, 60168, 60178, 60138, 60149, 60159, 60169, 60179, 60139, 60150, 60160, 60170, 60180,