WhichRecall
Class IIReported 2026-02-25

Brand Name: Spectrum IQ Product Name: Infusion Pump Model/Catalog Number: 3570009 Software Version: N/A Product Description: Infusion Pump, one unit per carton Component: N/A

Recalled by Baxter Healthcare Corporation

Reason for Recall

Pumps were released without full testing being performed, including occlusion alarm testing.

Advertisement
Recall Number
Z-1368-2026
Classification
Class II
Status
Ongoing
Category
device
Initiated
2026-01-13
Reported
2026-02-25
Distribution
States: Ohio Countries: United States
Code Info
Lot Code: GTIN 00085412610900, Serial Numbers 3815663, 3823468, 3824743, 3825234, and 3825599