WhichRecall
Class IIReported 2026-03-04

Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure

Recalled by Agiliti Health - Ellis

Reason for Recall

Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes

Advertisement

This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.

Recall Number
Z-1472-2026
Classification
Class II
Status
Ongoing
Category
device
Initiated
2026-01-21
Reported
2026-03-04
Distribution
US Nationwide distribution.
Code Info
UDI-DI 0084569904914; All serial numbers with a manufacturing date prior to 12/04/2025