Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1

Recalled by Medtronic Perfusion Systems

Reason for Recall

During the manufacturing, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned

Advertisement

The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.

Recall Number
Z-1473-2026
Classification
Class II
Status
Ongoing
Category
device
Initiated
2026-01-13
Reported
2026-03-04
Distribution
US Nationwide distribution in the states of FL and IN.
Code Info
UDI-DI: 00763000543693; Lot Number: 0232837110
Source: openFDA enforcement report. WhichRecall is not affiliated with the FDA. See our disclaimer.