WhichRecall
Class IIReported 2026-03-11

EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5

Recalled by Staar Surgical AG

Reason for Recall

Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.

Advertisement
Recall Number
Z-1480-2026
Classification
Class II
Status
Ongoing
Category
device
Initiated
2026-01-21
Reported
2026-03-11
Distribution
International distribution to the countries of India, Iran, Korea, Saudi Arabia.
Code Info
UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964, T2286971, T2286973, T2286979, T2286980, T2286981, T2286988
Source: openFDA enforcement report. WhichRecall is not affiliated with the FDA. See our disclaimer.