Class IIReported 2026-03-18

IBA Proton Therapy System - PROTEUS 235

Recalled by Ion Beam Applications S.A.

Reason for Recall

It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration

Advertisement

In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.

Recall Number
Z-1497-2026
Classification
Class II
Status
Ongoing
Category
device
Initiated
2026-02-03
Reported
2026-03-18
Distribution
Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.
Code Info
PROTEUS 235; UDI-DI: 05404013801138; Serial Numbers: SBF103 (JP), SBF124 (EU), SBF133 (US), SBF140 (EU)
Source: openFDA enforcement report. WhichRecall is not affiliated with the FDA. See our disclaimer.