Class IIReported 2026-03-18

DxC 700 AU, REF: B86444, B86446

Recalled by Beckman Coulter Mishima K.K.

Reason for Recall

A delay in results may occur

Advertisement

When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment.

Recall Number
Z-1552-2026
Classification
Class II
Status
Ongoing
Category
device
Initiated
2026-01-30
Reported
2026-03-18
Distribution
US Nationwide distribution including in the states of NE, WA, PA, CA, AZ, NC, TX, NV, CT, AL, NJ, TN, MT, ID, GA, SC, OH, KS, WI, HI, FL, VA, MI, NH, ME, NY, OR, LA, PR, MN, IL, MD, MO, MS, OK, AR.
Code Info
REF/UDI-DI/Serial Numbers: Newer Instrument Installed with PC preinstalled with Microsoft SQL Server 2022 Express: B86444/14987666542545/2025060063, 2025060073, 2025060074, 2025060075, 2025060079, 2025060085, 2025060086, 2025060087, 2025060091, 2025060098, 2025060099, 2025060124, 2025060125, 2025060126, 2025060127, 2025060128, 2025060129, 2025070139, 2025070156, 2025070159, 2025070160, 2025070161, 2025070164, 2025070167, 2025070168, 2025070169, 2025070178, 2025070180, 2025070181, 2025070182, 2025070184, 2025070186, 2025070191, 2025070192, 2025070193, 2025070194, 2025070204, 2025080208, 2025080217, 2025080219, 2025080236, 2025080237, 2025080238, 2025080241, 2025080242, 2025080246, 2025080247, 2025080250, 2025090266, 2025090267, 2025090270, 2025090271, 2025090273, 2025090274, 2025090275, 2025090281, 2025090283, 2025090285, 2025090286, 2025090287, 2025090289, 2025090290, 2025090291, 2025090292, 2025090293, 2025090294, 2025090295, 2025090296, 2025090297, 2025090298, 2025090299, 2025090301, 2025090302, 2025090303, 2025090304, 2025090306, 2025100370, 2025100371, 2025100372, 2025100373, 2025100421, 2025100422, 2025100423, 2025100424, 2025100425, 2025100426, 2025100427, 2025110428, 2025110429, 2025110437, 2025110442, 2025110445, 2025110452, 2025110454, 2025110456, 2025110457, 2025110458, 2025110462, 2025120464, 2025120468, 2025120469, 2025120471, 2025120475, 2025120477, 2025120478, 2025120479, 2025120480, 2025120482, 2025120483, 2025120484, 2025120485, 2025120486, 2025120487, 2025120488, 2025120489, 2025120490, 2025120492, 2025120493, 2025120494, 2025120495, 2025120496. B86446/14987666542514/2025090277, 2025090284. Older Instrument Replaced with PC preinstalled with Microsoft SQL Server 2022 Express: B86444/14987666542545/2017050064, 2017100174, 2018030312, 2018040352, 2018100561, 2018110599, 2019010712, 2019010713, 2019111170, 2021021852, 2021102318, 2021122458, 2022042631, 2022052705, 2022062775, 2023063701, 2024064408, 2024064419, 2024064440
Source: openFDA enforcement report. WhichRecall is not affiliated with the FDA. See our disclaimer.
Advertisement