WhichRecall
Class IIReported 2026-04-01

Custom Abutment Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 027.4620;

Recalled by Straumann USA LLC

Reason for Recall

A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments

Advertisement

More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.

Recall Number
Z-1632-2026
Classification
Class II
Status
Ongoing
Category
device
Initiated
2026-03-17
Reported
2026-04-01
Distribution
Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, MN, MO, MT, OR, WA; and the country of Canada.
Code Info
Catalog Number/Art.: 027.4620; UDI-DI: 07630031713766; Order number (lot): 92465US_12039_00_b00 , 92465US_12040_00_b00 , 92465US_12047_00_b00 , 92465US_12048_00_b00 , 70959US_02227_00_b00 , 94021US_02348_00_b00, 70416US_07543_00_b00 , 91395US_01687_00_b00 , 91545US_03284_00_b00 , 91545US_03285_00_b00 , 71711US_00420_00_b00 , 91853US_06313_00_b00 , 72747US_00247_00_b00;
Source: openFDA enforcement report. WhichRecall is not affiliated with the FDA. See our disclaimer.
Advertisement