Class IIReported 2026-04-01

Brand Name: Olympus PK Cutting Forceps, 5mm 33cm Product Name: PK Cutting Forceps Model/Catalog Number: PK-CF0533 Product Description: An assembly of devices intended to generate and apply radio-frequency (RF) alternating current to soft tissues for cutting and coagulation during an endoscopic or open surgical procedure.

Recalled by Olympus Corporation of the Americas

Reason for Recall

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process

Advertisement

Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Recall Number
Z-1640-2026
Classification
Class II
Status
Ongoing
Category
device
Initiated
2026-02-25
Reported
2026-04-01
Distribution
Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
Code Info
Model: PK-CF0533; UDI: 00821925035867; Lot Numbers: All unexpired;