Elevated endotoxin levels identified in affected lots. Issue may elicit strong inflammatory response if used, including local tissue irritation, delayed wound healing, or systemic effects such as fever and hypotension. In severe cases, septic shock may result.
The instructions for use (IFUs) on specified Cardinal Health Chest Drainage Units (CDUs) and accessories have been updated to clarify the intended target patient population as adults, 18 years and older.
Fabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If delaminated product is used and surgical strikethrough occurs, there may be risk of infection.
Fabric may experience premature delamination when using included laundering instructions. Identifying delaminated gown or drape in a surgical setting may result in brief delay in patient care. If delaminated product is used and surgical strikethrough occurs, there may be risk of infection.
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.
Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.
Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
A smaller open suction catheter, the 8 Fr Suction Catheter, may have been packed within the kit instead of the required 14 Fr Suction Catheter. Use of the smaller size catheter may result in hypoxia and/or respiratory failure related to partial or full airway obstruction. Airway obstruction may require immediate emergent medical intervention to prevent respiratory failure or death. Other potential outcomes include prolonged procedure, airway irritation, or prolonged dyspnea.
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
A smaller open suction catheter, the 8 Fr Suction Catheter, may have been packed within the kit instead of the required 14 Fr Suction Catheter. Use of the smaller size catheter may result in hypoxia and/or respiratory failure related to partial or full airway obstruction. Airway obstruction may require immediate emergent medical intervention to prevent respiratory failure or death. Other potential outcomes include prolonged procedure, airway irritation, or prolonged dyspnea.
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Firm received multiple complaints regarding tubing disconnecting from oxygen mask. If issue occurs during use, may lead to delay in patient care and patient may experience shortness of breath leading to hypoxia, which may require medical intervention to prevent further respiratory deterioration.
The items may have weak packaging seals that may result in a breach of sterility if the seal fails and may lead to an increased risk of infection if product is utilized.
The items may have weak packaging seals that may result in a breach of sterility if the seal fails and may lead to an increased risk of infection if product is utilized.
Medline has identified that all lots of the Curad Touch-Free and Quick-Strip Bandages may exhibit open seals in the packaging which may compromise the sterility of the product.
Medline has identified that all lots of the Curad Touch-Free and Quick-Strip Bandages may exhibit open seals in the packaging which may compromise the sterility of the product.
Medline has identified that all lots of the Curad Touch-Free and Quick-Strip Bandages may exhibit open seals in the packaging which may compromise the sterility of the product.
Medline has identified that all lots of the Curad Touch-Free and Quick-Strip Bandages may exhibit open seals in the packaging which may compromise the sterility of the product.
Medline has identified that all lots of the Curad Touch-Free and Quick-Strip Bandages may exhibit open seals in the packaging which may compromise the sterility of the product.
Medline has identified that all lots of the Curad Touch-Free and Quick-Strip Bandages may exhibit open seals in the packaging which may compromise the sterility of the product.
There have been multiple complaints of drill bit fusing to the bushing/sleeve within the drill guide during use. Retrieving a replacement drill bit and drill guide may prolong the surgical procedure.
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing BD Luer Tip Caps. BD has issued a recall of their Luer Tip Caps after confirming through internal testing that certain BD Luer Tip Cap trays failed routine biocompatibility testing. Although the tip cap covers a small surface area of the syringe and does not come in direct contact with patients, there is the potential that the failure of a biocompatibility test may have varying consequences ranging from no harm or clinical impact to potential for sensitization, irritation, pyrogenicity, acute systemic toxicity or hemolysis depending on the nature and duration of contact with the syringe.
Medline has identified that all lots of the Curad Touch-Free and Quick-Strip Bandages may exhibit open seals in the packaging which may compromise the sterility of the product.